Description
Features
|
Model |
HI960 |
HI960RS |
|
Measuring item |
Ion / Temp. |
|
|
Range |
0.00 ~ 20000 ppm |
|
|
-2000.0 ~ 2000.0 mV |
||
|
Resolution |
1 / 0.1 / 0.01 ppm |
|
|
0.1 mV |
||
|
Accurracy |
±0.06% ±1 digit |
|
|
Temp. |
-10.0~110.0 ℃ |
|
|
Temp. accuracty |
±0.2 °C ±1digit |
|
|
Temp. compensation |
Pt1000 / NTC30K |
|
|
Remark |
2 ~3 point or offset calibration |
|
|
Output |
Analog output 1 : Isolated DC 0/4~20mA corresponding to measurement max. load 500Ω |
|
|
Communication |
x |
RS485 |
|
Contact output |
Contact output : Relay ON/OFF, max. 240VAC 0.5A |
|
|
Washing setting |
ON : 0~9999 seconds / OFF : 0~999.9 hours |
|
|
Water proof |
IP 65 |
|
|
Temperature for use |
0 ~ 50 ℃ |
|
|
Dimension |
96 x 96 x 145 mm / 0.4 kg |
|
|
Power supply |
100 ~ 240VAC / ±10%, 50/60Hz |
|



![[NeoMet - K series ] pH/Ion/DO/Conductivity/Temp Meter (4 Channel/ Customised ) K9000-M](https://wisdomlead.ca/wp-content/uploads/2025/05/GOODS2_1447050653-min-300x300.png)